Sun Microsystems, Inc.  Sun System Handbook - ISO 3.4 June 2011 Internal/Partner Edition
   Home | Current Systems | Former STK Products | EOL Systems | Components | General Info | Search | Feedback

Asset ID: 1-71-1013594.1
Update Date:2011-02-09
Keywords:

Solution Type  Technical Instruction Sure

Solution  1013594.1 :   SL8500 - Calibration of Operator Panel Screen  


Related Items
  • Sun StorageTek SL8500 Modular Library System
  •  
  • Sun StorageTek SL8500 Modular Library System
  •  
Related Categories
  • GCS>Sun Microsystems>Storage - Tape>Libraries - SL-Series
  •  

PreviouslyPublishedAs
219143


Oracle Confidential (PARTNER). Do not distribute to customers
Reason: Confidential for Partners and Oracle Support personnel

Applies to:

Sun StorageTek SL8500 Modular Library System
Sun StorageTek SL8500 Modular Library System - Version: Not Applicable and later    [Release: N/A and later]
All Platforms
Checked for relevance on 9-Feb-2011.

Goal

Calibration of Operator Panel Screen.

Solution

Steps to Follow
The following information is found in "SL8500 Replaceable Parts Catalog, PN96139, Rev-E", page 343.

 

Calibration - Hardware Method

If the operator panel requires calibration:

1. Power-off the operator panel.
2. Power-on the operator panel.
3. Carefully watch the screen. When the Windows desktop first appears, touch the screen; then, as soon as possible, close (click the "X" in the upper right hand corner) of the black DOS screen.

Note: If you do not close the DOS screen, the StreamLine Library Console application will activate within five seconds.

4. Look to the bottom right for a PM icon, touch this icon.
                A menu will appear.
5. Select Control Panel.
6. Select the Calibrate tab.
7. When the window appears, select Calibrate again.
                At the top of the blank screen, you will see the words Point here (next to a small circle).
8. Touch the exact middle of the circle with the stylus.
9. Continue touching the exact middle of the circle as it moves to each location on the screen.
10.After all four screen/circle locations have been calibrated, click OK.
11. To initialize the StreamLine Library Console:
a. Click Start in the lower left corner.
b. Select Startup from the menu.
c. Click SLConsole.

Calibration - Firmware Method

1. At the screen, touch START -> Programs -> Penmount Utilities 2000 -> Penmount Control Panel

2. Select Calibrate

3. Perform steps 7 through 11 (above).

Return to "Operator Panel."




SL8500, Screen, How to Calibrate Op Panel Screen, Operator panel, Calibration, Control panel, Calibrate, Op Pane
Previously Published As STKKB80297

Attachments
This solution has no attachment
  Copyright © 2011 Sun Microsystems, Inc.  All rights reserved.
 Feedback